2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 425
CO- 10.0.44 PRE DEV VALUE LESS THAN 135 / CONDUCTIVITY STABILITY
a) Post Dev val (right column) is the conductivity difference between now and one (1)
minute ago from Post-Conductivity Cell #13. If no alarms it changes continually!
b) After no more than three (3) minutes, if not already, Post Dev val should fall to less than 70. TWO (2)
possible scenarios:
1) IF Post Dev val is OR falls to less than 70: See procedure number CO- 10.0.46 (page 426).
2) IF Post Dev val NEVER falls below 70: Post Cond is unstable! ENSURING no alarms occurred
see parts a and b below:
a) Per the Figure above, using a flashlight and WITHOUT LOOKING AWAY, for one minute, if air
bubbles are flowing through the Post Dialyzer Cond Cell #13’s locate and repair the air leak!
b) If (and ONLY if) absolutely sure of no air proceed to page 429, procedure number
CO- 10.0.60.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals