2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 490
TMP- 2.0.1 POSITIVE PRESSURE LEAK CHECKS
Small external leaks cause BIG problems! For the following USE A FLASHLIGHT! FEEL AROUND!
NOTE! Not being THROUGH may cost unnecessary troubleshooting time!
a) THREE (3) external leak checks:
CHECK #1: Figure right, the ENTIRE LENGTH of
the dialyzer lines and ESPECIALLY the
Dialyzer connectors!
CHECK #2: The ENTIRE hydraulic compartment!
CHECK #3: Figure right, ENSURE no leaks from inside the DiaSafe
®
Filter housing and its tubing
b) Figure below, any water seen coming through Diasafe
®
Filter #92*?
* Water staying inside the filter is NORMAL!
Yes Valve #28 is leaking! Proceed to
page 516, procedure number
TMP- 9.0.0.
No fluid seen! See procedure number TMP- 2.0.2 (page 491).
Figure 68 – DiaSafe
®
Filter / Test Valve #28 / Filter #92