EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #706 background imageLoading...
Page #706 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 702
P- 2.0.5 ISOLATE POWER LOGIC CABLE
a) CAUTION! Signals are about to be measured at pins that are VERY close to others and touching
them together with a standard meter lead MAY CAUSE massive damage! As directed below, make
your RED meter lead PROTECTED! DO NOT CONTINUE UNTIL YOU HAVE DONE
THIS!
b)
Per the Figure below, at the top of the Power Logic Board, closest to the screen, locate its 20-pin
‘X2’ ribbon cable.
c) Measure at the rear (solder) side of the X2 connector at
pin 1 (bottom row, first pin from the REAR of
the machine). More than 10.0 volts DC?
Yes More than 10 volts! See procedure number P- 2.0.6 (page 703).
No Less than 10 volts! Replace the 20-pin Power Logic X2 ribbon cable.
Figure 130 – Electronic Card Cage / Power Logic Board / Pin1

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals