2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 418
CO- 10.0.20 PROBLEM WITH POST DIALYZER TEMP / ISOLATE TEMP SENSOR #44
a) Figure right, from the distribution board, unplug the
11
th
Connector cap from the left. This is
Temperature Sensor NTC #44, x44, NTC-POST”!
b) Screen 0’s Post Dialyzer Temp MUST be LESS THAN
34.0° C at this time! If NOT, ENSURE Post Dialyzer
Temperature Sensor #44 was unplugged!
c) Using a flashlight, check the vacant x44position. If
corrosion or damage is located this may be the
problem!
d) Figure right, install the
6.04 KΩ plug, from the TWO-RESISTOR SET,
into the Post Dialyzer Temp Sensor’s #44 distribution board position, “x44,
NTC-POST”.
e) Is screen 0’s Post Dialyzer Temp between 35.5 and 38.5° C?
Yes Post Temp between 35.5 and 38.5! Proceed to page 420, procedure number CO- 10.0.26.
No Post Temp is NOT between 35.5 and 38.5! See parts a and b below:
a) ENSURE the 6.04 KΩ plug from the TWO-RESISTOR SET is placed properly at
position “x44, NTC-POST”. If NOT, repeat (ABOVE) procedure number CO- 10.0.20
(page 418) AND / OR CONSIDER using the 6.04 KΩ plug from another TWO-RESISTOR
SET.
b) If SURE the plug is placed correctly see procedure number CO- 10.0.22 (page 418).
CO- 10.0.22 POST DIALYZER TEMP IS NOT BETWEEN 35 AND 38°C
a) ENSURE connector labeled #7 (Pre Cond Cell #7) is in position X7, COND”.
b) Enter Service Mode → Calibrate Sensors → Post Temp Sensor. NOTE! This is a non-routine
troubleshooting procedure! Follow the instructions carefully to avoid error!
c) Press ‘Enter’. The screen says 1. Connect a 6.808 K ohm resistor….
d) Per the Figure right, place the
34° C (6.808 KΩ) plug, from the FOUR-
RESISTOR SET into distribution board position, “x44, NTC-POST”.
e) Is the screens [Post-Temperature Reference] more than 20?

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals