2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 85
SPENT SIDE THEORY: Per the Figure below, when “Select Program” is up the top balancing chamber
(BC) valves (V31 V34) are supposed to be open, the bottom BC valves (V35 V38) are closed. Flow
Pump #21 is off which is why it must be plugged into the Deaeration Pump’s distribution board position.
Also, the red dialyzer line is placed into a bucket of water to provide the Flow Pump with a source of water.
PART 3 THEORY: V36 is plugged into V33’s position which should open V36. If both V36 and V32 are
open the Flow Pump draws from the bucket and sends it through the spent side of the balancing chamber,
through V30, to drain where it is measured.
PART 4 THEORY: V38 is plugged into V31’s position which should open V38. If both V31 and V38 are
open the Flow Pump draws from the bucket and sends it through the spent side of the balancing chamber
to drain.
Figure 18 – Parts 3 and 4
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals