2008T Troubleshooting Guide
P/N 490292 Rev. A
CO- 1.0.3.7 NOT EQUIPPED WITH BIBAG / CONDUCTIVITY LOW
A) Figure right, while watching through the inlet tubing,
disconnect the ACID connector from the jug, for seven (7)
seconds THEN FIRMLY reconnect!
B) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1) IF (and ONLY if) air / acid is moving in one
direction ONLY, towards the machine! See
part C.
2) IF air / acid is NOT moving OR moving back and
forth! See procedure number CO- 4.1.0 (page 375).
C) Figure right, watching through the inlet tubing,
disconnect the BICARB connector from the jug, for
seven (7) seconds THEN FIRMLY reconnect!
D) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1) IF (and ONLY if) air / bicarb is moving, in one
direction ONLY, towards the machine! See
procedure number CO- 1.0.3.8 (page 351).
2) IF air / bicarb is NOT moving OR moving back and forth! See procedure number CO- 4.1.0
(page 375).