2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 350
CO- 1.0.3.7 NOT EQUIPPED WITH BIBAG / CONDUCTIVITY LOW
A) Figure right, while watching through the inlet tubing,
disconnect the ACID connector from the jug, for seven (7)
seconds THEN FIRMLY reconnect!
B) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1) IF (and ONLY if) air / acid is moving in one
direction ONLY, towards the machine! See
part C.
2) IF air / acid is NOT moving OR moving back and
forth! See procedure number CO- 4.1.0 (page 375).
C) Figure right, watching through the inlet tubing,
disconnect the BICARB connector from the jug, for
seven (7) seconds THEN FIRMLY reconnect!
D) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1) IF (and ONLY if) air / bicarb is moving, in one
direction ONLY, towards the machine! See
procedure number CO- 1.0.3.8 (page 351).
2) IF air / bicarb is NOT moving OR moving back and forth! See procedure number CO- 4.1.0
(page 375).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals