2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 110
F- 9.0.6 ACFS = 0
a) Return CFS #10’s connector to distribution board position “x10, CFS”. If returned properly ACFS =
between 3.0 and 8.1.
b) From the Home screen, ENSURE Dialysate Flow is set at
500 ml/min.
c) Is the TMP window RED?
Yes Red TMP window! A) Press and release the ‘Reset’ key then immediately press and hold it
for three seconds; B) Allow thirty (30) seconds; C) If a TMP alarm reoccurs attempt RESET
up to twice more BEFORE continuing to procedure number F- 9.0.7 (page 110).
No TMP window is pale yellow/white! See procedure number F- 9.0.7 (page 110).
F- 9.0.7 ANALYZE TMP
TMP is STABLE if (and ONLY if) the TMP Window REMAINS white AND the TMP does NOT change more
than +/- 60 mmHg in three (3) minutes. TWO (2) possible scenarios:
1) IF (and ONLY if) TMP is STABLE: See procedure number F- 9.0.8 (page 110).
2) IF TMP is UNSTABLE: A procedure, in a different Section, is performed next. IMPORTANT! NOTE
this page and procedure number (F- 9.0.7) as you may prompted to return to here: Perform parts A
and B below:
A) BEFORE continuing to part B, proceed to
page 572, SECTION 18A DIAGNOSTIC VALVE
LEAK TESTS.
B) If a balancing chamber valve leak was not located in part A, see procedure number F- 9.0.8
(page 110).
F- 9.0.8 BALANCING CHAMBER DIAPHRAGM TEST
A procedure, in a different Section, is performed next. IMPORTANT! NOTE this page and procedure
number (F- 9.0.8) as you may prompted to return to here:
a) BEFORE continuing to part b, proceed to
page 580, to perform SECTION 19 - TESTING FOR A
LEAKING BALANCING CHAMBER DIAPHRAGM.
b) If a leaking diaphragm is not located in part a, return to Dialysis Program (“Select Program” → ‘Dialysis’
→ ‘Enter’)!
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘Enter’
d) If BC Switch (debug screen 6) = 897 (constantly) AND ACFS (debug screen 10) remains between 3.0
and 8.0. TWO (2) possible bad components: 1) CFS transducer #10
a
OR; 2) Sensor board
b,c
.
a
To LOCATE CFS transducer #10 refer to Figure 21 (page 100);
b
To LOCATE the Sensor board see
Figure 4A (page 10).
c
To prevent “Cond Offset Failure” place the machine into T and C Mode (refer to OPERATING
MODES page 19)).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals