2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 610
1) IF (and ONLY if), during this procedure, tubing ‘blew off’ the DiaSafe
®
filter or the
top side Balancing Chamber Valves (see Figure below): A) Turn the machine off;
B) Reattach and tie wrap all tubing; c) Repeat (ABOVE) procedure number
LEAKING- 5.1.0 (page 609).
2) IF tubing DID NOT ‘blow off’: A) Reconnect the drain tubing; B) NOTE
ONLY VALVE
#25 will be checked and proceed to page 210, TROUBLESHOOTING A VALVE.
LEAKING 5.2.0 VALVE #25 OKAY / ISOLATE VALVE #26
a) Press ‘Esc’ then ‘Enter’ twice to call the “Select Program” screen!
b) Remove the red dialyzer connector from the shunt door and place it on the floor!
c) Close the door!
d) Per the Figure above, clamp and remove Valve #26’s
INPUT tubing!
e) Fill the 60 ml syringe with water then attach it to Valve #26’s INPUT nozzle.
f) Push AS HARD AS YOU CAN on the syringe plunger. Can you push ALL 60 ml through Valve #26 within
FIVE (5) seconds?
Yes Valve #26 is okay! Return the dialyzer connector to the shunt then see procedure number
LEAKING- 5.3.0 (page 611).
No Significant resistance felt! NOTE
ONLY VALVE #26 will be checked and proceed to
page 210, TROUBLESHOOTING A VALVE.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals