2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 206
Yes Air moving ALL THE WAY through Valve #29! See procedure number CLEAN- 6.1.0
(page 206).
No
Air is NOT moving through Valve #29! See parts A AND b below:
A) Remove the clamp from Valve #30!
B) The Troubleshooting Guide cannot locate an immediate problem. Return to Dialysis
Program and troubleshoot potential hydraulic problems from there!
CLEAN- 6.1.0 ‘AIR’ MOVING THROUGH VALVE #29
a) Turn the machine OFF and LEAVE IT OFF!
b) Screw a 60 ml syringe, COMPLETELY FILLED WITH
AIR, onto the Fluid Sample Port.
c) Push hard on the syringe plunger while watching Valve
#29’s OUTPUT tubing. Do you see air moving through
Valve #29?
Yes Air moving through Valve #29! Valve #29 is
bad (sticking open).
No Air IS NOT moving through Valve #29! The Actuator-Test Board is bad!
LEFT BLANK INTENTIONALLY