2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 234
b) Place the machine in RINSE!
c) In RINSE, is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes ‘Bob’ moving! See procedure number T- 1.0.2 (page 234).
No In RINSE, bob’ is NOT moving! See parts a THROUGH c below:
a) Connect to acid and LIQUID bicarb.
b) RETURN TO DIALYSIS PROGRAM (“Select Program” → ‘Dialysis’ → ‘Enter’)!
c) Proceed to page 23, SECTION 1 - FLOW ERRORS IN DIALYSIS
PROGRAM
T- 1.0.2 ‘BOB’ MOVING
A) Press ‘Esc’ then ‘Enter’ twice to call Select Program” but DO NOT press
‘Dialysis’ till instructed!
B) Figure right, TWO (2) possible scenarios:
1) IF (and ONLY if) NOT equipped with a
bibag Connector: See procedure
number T- 1.0.3 (page 235).
2) IF equipped with a bibag Connector:
See parts a AND b below:
a) A bibag disposable CANNOT be
attached. Instead connect to a jug of
known good liquid bicarb!
b) With “Select Program” remaining up, see procedure number T- 1.0.3 (page 235).
NOTE! Temperature problems may cause the bibag system to seemingly malfunction.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals