2008T Troubleshooting Guide
P/N 490292 Rev. A
b) Place the machine in RINSE!
c) In RINSE, is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes ‘Bob’ moving! See procedure number T- 1.0.2 (page 234).
No In RINSE, ‘bob’ is NOT moving! See parts a THROUGH c below:
a) Connect to acid and LIQUID bicarb.
b) RETURN TO DIALYSIS PROGRAM (“Select Program” → ‘Dialysis’ → ‘Enter’)!
c) Proceed to page 23, SECTION 1 - FLOW ERRORS IN DIALYSIS
PROGRAM
T- 1.0.2 ‘BOB’ MOVING
A) Press ‘Esc’ then ‘Enter’ twice to call “Select Program” but DO NOT press
‘Dialysis’ till instructed!
B) Figure right, TWO (2) possible scenarios:
1) IF (and ONLY if) NOT equipped with a
bibag Connector: See procedure
number T- 1.0.3 (page 235).
2) IF equipped with a bibag Connector:
See parts a AND b below:
a) A bibag disposable CANNOT be
attached. Instead connect to a jug of
known good liquid bicarb!
b) With “Select Program” remaining up, see procedure number T- 1.0.3 (page 235).
NOTE! Temperature problems may cause the bibag system to seemingly malfunction.