2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 65
F- 6.1.8 VERIFY FLOW RATE (BC SWITCH)
a) Do NOT reset alarms!
b) Call debug screen 6. WITHOUT LOOKING AWAY, watch BC Switch (middle
column) for two (2) FULL minutes OR until it EVER = 897 or more. FOUR (4)
possible scenarios 1) or 2) or 3) or 4) below:
1) IF (and ONLY if) BC Switch REMAINS ALWAYS = 897: ENSURING a “No Water” alarm has
NEVER occurred proceed to page 97, procedure number F- 8.0.0.
2) IF (and ONLY if) BC Switch ‘cycles’ to more than four-hundred (400) for two (2) seconds
then to less than 3.0 for nine (9) seconds: Proceed to
page 70, procedure number F- 7.0.0.
3) IF (and ONLY if) BC Switchcyclesto more than four-hundred (400) but NEVER to 897:
See procedure number F- 6.1.9 (page 65).
4) IF BC Switch is NEVER more than four-hundred (400): See parts a THROUGH c below:
a) Call debug screen 1. If NPHT (upper right) is decreasing press and hold
the keyboard’s “1” key until it is more than 600 then see part b If (and
ONLY if) NPHT = 720 skip to part c.
b)
Allow forty-five (45) seconds BEFORE continuing to part c!
c) Return to debug screen 6 and continue to watch BC Switch for two (2) minutes. It may change
slightly but should REMAIN ALWAYS between 203 and 248?
Yes BC Switch is REMAINING between 203 and 248! The Flow Error is not
presenting at this time. It may be extremely intermittent! Return to
page 52,
procedure number F- 3.8.0
No BC Switch is NOT remaining between 203 and 248! Call the Home screen. TWO
(2) possible scenarios below:
1) IF (and ONLY if) [Dialysate Flow] has remained at 800 ml/min: See
procedure number F- 6.1.9 (page 65).
2) IF [Dialysate Flow] has NOT remained 800 ml/min: Set it to 800, allow thirty
(30) seconds, then return to (ABOVE) procedure number F- 6.1.8 (page 65).
F- 6.1.9 DETERMINE FLOW ERROR
From the Home screen, is [Conductivity] more than 14.8 mS?
Yes [Conductivity] = More than 14.8 mS! Proceed to
page 70, procedure number F- 7.0.0.
No [Conductivity] is less than 14.8 mS! Proceed to
page 69, procedure number F- 6.5.0.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals