2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 197
CLEAN- 4.4.2 MORE THAN 300 ML / VALVE #24 OPEN / ISOLATE VALVE #25
a) With the dialyzer connector out of the shunt ENSURE the external flow indicator’s ‘bob’ is rising at least
¼ way up in the sight tube!
b) Return the connector to the shunt and close the door.
c) Does the external flow indicator’s ‘bob’ now rise at least ¼ way up?
Yes ‘Bob’ moving now! See procedure number CLEAN- 4.6.0 (page 198).
No ‘Bob’ NOT moving! See parts a AND b below!
a) ENSURE no restrictions in the ‘from dialyzer (red) line’ especially the external filter!
b) If no restrictions, NOTE
ONLY VALVE #25 will be checked and proceed to page 210,
TROUBLESHOOTING A VALVE
LEFT BLANK INTENTIONALLY