2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 79
F- 7.0.5 ISOLATE VALVE #26 / VALVE #25 ACTUATOR-TEST BOARD CONTROL
a) ENSURE “Select Program” REMAINS up.
b) EXCEPT for valve #37 and valve #30, ENSURE ALL OTHER valves are plugged into their PROPER
distribution board positions!
c) TWO (2) ‘touch’ checks on valve #26 AND #25’s (black) solenoids:
8TOUCH CHECK #1: Per Figure 15 below touch Valve #26’s solenoid.
TOUCH CHECK #2: Per
Figure 16 (next page), above (blue) Dialysate Pressure Transducer #9, is
Valve #25. Touch its solenoid then see part d.
d) Per Table J below, respond according to BOTH ‘touch’ checks:
Table J Valve #26 / #25
Valve #26’s
Solenoid
Valve #25’s
Solenoid
Your Response
Warm (or Hot) Warm (or Hot) See procedure number F- 7.0.6 (page 80)
Cold Does not matter
NOTE
ONLY VALVE #26 will be checked and proceed to
page 210, TROUBLESHOOTING A VALVE
Does not matter Cold
NOTE
ONLY VALVE #25 will be checked and proceed to
page 210, TROUBLESHOOTING A VALVE
Figure 15 – Valve #26

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals