2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 337
CO- 1.0.0 ISOLATE DIALYSATE SETTINGS
A) At the bottom of the screen, press the ‘Dialysate’ tab.
B) Figure below, ENSURE the ‘Selected Acid’ matches the acid connected to the machine.
Figure 55 – Dialysate Screen
C) Set [Base Na+] to 137 and press ‘Enter’.
D) Set [Bicarbonate] to 33 and press ‘Enter’.
E) From the Home screen, ENSURE [Dialysate Flow] is on to at 500 ml/min or more.
F) If the ‘Selected Acid’ AND / OR [Base Na+] AND / OR [Bicarbonate] was changed, allow five (5)
minutes BEFORE continuing to part G.
G) Based on [Conductivity] now, THREE (3) possible scenarios:
1) IF (and ONLY if) LESS THAN 13.0 mS: See procedure number CO- 1.0.1 (page 338).
2) IF (and ONLY if) STABLE between 13.0 and 14.3: Proceed to
page 363, procedure number
CO- 1.0.9.
3) ALL OTHERS: Proceed to
page 353, procedure number CO- 1.0.4.